Summary
Sy-9900 fixed pollution source on-line monitoring system is an ultra-low emission CEMS system launched by our company in combination with the emission standards of various industries in
In the particle monitoring part, the high-temperature extraction laser front scattering principle technology is adopted, and the interference of wet flue gas condensate and water drop on particle monitoring is effectively overcome by heating sampling method.
Product features
Heating type sampling probe rod to reduce the loss of gas to be measured in the sampling process.
The pretreatment adopts acid condenser to inhibit the loss of SO2 in the process of sample gas condensation, and the loss rate is less than 3mg / m3.
No and NO2 can be measured directly without converter.
The analyzer adopts small range, high-precision AMETEK ultraviolet absorption gas analysis.
Fast bypass, short system delay time.
The system has few moving parts, stable and reliable operation, and extremely low failure rate.
紫外分析仪SO2 MONO2O2
高温抽取式粉尘仪
技术参数
气体分析仪 | |
测量原理 | 双光源多波长高分辨率不色散紫外吸收技术 |
分析组分 | SO2, NO, NO2, NOx, O2(电化学) |
测量范围 | SO2:0-75mg/m³, NO:0-100mg/m³;NO2:0-100mg/m³ O2:0-25% |
线性偏差 | ≤1%FS |
漂移 | ≤1%FS/周 |
重复性 | ≤0.5%FS |
响应时间 | T90≤30S |
环境温度 | 0-50 ˚C |
供电电压 | AC230V 50HZ, 200VA |
测量输出 | 4-20mA, Modbus 485, |
重量 | 18.2Kg |